What does this change, and why?

A finding from the contact work, recorded in contact_in_run.md (section 6) with a question for the FEBio team in deviations.md. Notes only; no code.

Sorting the active contact pairs of the open mitral input at full closure by the distance between their two points in the reference configuration:

Reference distance Active pairs Contact area of the points they touch
Below 0.75: nearby points of one leaflet 268 17.5
0.75 to 3 6 0.3
Above 3: points of opposing leaflets 3,146 72.9

With the normal test w_tol, about a fifth of the reported contact area (90.7 here, 89.6 in the stored FEBio run) comes from pairs between nearby points of the same leaflet, presumably at creases, and not from one leaflet meeting another. Excluding pairs by reference distance in place of the normal test, as the original formulation does for self-contact, removes exactly those pairs and leaves the rest: its contact area is 73.8. That closes the open reading under decision K6 for the smooth contact tier, and raises a question about the contact area as a measure of coaptation.

Merged by the author; comments afterwards are welcome.

Validation

  • Tests added or changed: none.
  • FEBio reference (input file, quantity compared, tolerance, FEBio version), or N/A: the stored FEBio 4.9.0 contact area of mitral_p2prolapse.feb, as in pull request 14. The pair counts are from a run of this code on a Quadro RTX 8000 with the optional cudss solve; the script is not in the repository.
  • Devices the full tier ran on (CPU, and the GPU model if CUDA; say so clearly if no GPU was used): none; documentation only. uvx pre-commit run --all-files and the documentation link check pass.

🤖 Generated with Claude Code

…on of a leaflet

Sorting the active contact pairs of the open mitral input at closure
by their distance in the reference configuration: 268 of 3420 are
between nearby points of the same leaflet (below 0.75), and the points
they touch make up 17.5 of the contact area of 90.7; the pairs between
opposing leaflets (above 3) account for 72.9, which is the contact
area of the variant that excludes pairs by reference distance in place
of the normal test. So that variant loses nothing that belongs to
closure (decision K6), and the contact area reported with w_tol
overstates coaptation by about a fifth on this valve. Recorded in the
contact note, with a question for the FEBio team in the deviations
note.

Co-Authored-By: Claude Fable 5.2 <noreply@anthropic.com>
Signed-off-by: Matthew Jolley <12954190+mattjolley@users.noreply.github.com>
@mattjolley
mattjolley merged commit 0acce84 into main Nov 28, 2026
2 checks passed
@mattjolley
mattjolley deleted the mattjolley/contact-area-finding branch November 28, 2026 16:17
Sign up for free to join this conversation on GitHub. Already have an account? Sign in to comment

Labels

None yet

Projects

None yet

Development

Successfully merging this pull request may close these issues.

1 participant